Oxylator® FR 300


The OXYLATOR® FR-300 should be used only by individuals who have adequate training in CPR techniques and the operation of oxygen powered resuscitators.





Description

The OXYLATOR® FR-300 is an oxygen powered resuscitator which requires no power source (i.e. battery) other than the supply of compressed oxygen in an appropriate cylinder or wall outlet. The oxygen supply hose of the OXYLATOR ® FR-300 should be connected to its pressurized source supplying a minimum constant flow of 40 litres per minute and between 45 psig to 80 psig or 3.0 bar to 5.5 bar. During the continuous cycling mode, the inspiratory flow of oxygen will be initiated upon complete, continuous depression of the OXYGEN RELEASE BUTTON. A sensing chamber in the OXYLATOR ® FR-300 allows the airway pressure to increase to the maximum limit of 20 cmH2O, at which point the system shuts off the inspiratory flow, allowing passive exhalation to take place. Upon sensing the completion of the exhalation phase, the valve automatically resets and allows the inspiratory flow to start a new cycle. Cycling continues as long as required, provided the OXYLATOR ® FR-300 is connected to pressurized oxygen and its OXYGEN RELEASE BUTTON is continuously, fully depressed, or engaged for continuous cycling mode.
Features

- Meets American Heart Association recommendations for
   "Adjuncts for Oxygenation, Ventilation, and Airway Control"
   (Part III, Adult Advanced Cardiac Life Support of Guidelines
   for Cardiopulmonary Resuscitation and Emergency Cardiac
   Care, 1992)

- Extremely light-weight, impact resistant and durable
- Simple to learn
- Easy to use
- Single use, quick change filter element
- Few moving parts, minimal maintenance required
- Airway pressure limit of 20 cmH2O
- Constant maximum flow rate of 30 litres per minute
- Manual and continuous cycling modes
- Compatible with oxygen supply tanks and regulators supplying
    between 45 and 80 psig inlet pressure and constant minimum
    flow rate of 100 litres per minute
- Hold-down oxygen release mechanism
- PEEP of 2- 4 cm H2O during continuous cycling mode
- Eliminates the risk of "air trapping" or "stacked breath"
- FIO2 of 1.0 during resuscitation
- Oxygen powered No waste of O2 supply for operation
- Audible and visual indication of airway obstruction
Specifications of Oxylator® FR-300

   

Model

Oxylator FR-300

Weight

180 g or 0,40 lbs

Dimensions

48 mm x 100 mm or 1,87 x 3,94 in

Housing material

Acetal

Required source pressure

3,0 – 5,5 bar or 45-80 psi

Required flow source

Minimum 40 L/min.

Dead space

20 ml

Inspiratoiry flow rate (constant flow)

Maximum 30 L/min.

Average time of oxygen supply
(auto mode)

Ca. 12 L/min.

Average inspiration/expiration ratio

1 – 1,5 of manually controlled

PEEP (in automatic mode)

2-4 cm H2O

Ventilatory frquency

Auto-adjusting, to lung capacity and compliance in automatic mode

Pressure relief

20 cm H2O (14,5 mm Hg)

Expiratory resistance

5 cm H2O in manual mode

Suitable body mass range

Greater than 10 kg or 22 lbs

Oxygen concentration (FiO2)

100% in automatic mode and connected with oxygen source of 100% oxygen

Filter

Single use, viral / bacterial

Mask/airway connection

External 22 mm; internal 15 mm

Usage temperature range

-30 °C tot 60 °C or -22 F to 140 F

Storage temperature range

-40 °C tot 70 °C or -40F to 158 F

Obstructed airway warning

Rapid cycling, audible and visual

Oxygen inlet connection

DISS, ISO Standards

Warranty

5 year, maintenance free warranty

*Technical specifications may vary slightly depending on the model and development

Copyright © 2017 www.oxylator.biz